多通道定標平均器是什麼 ?
多通道定標平均器是用來累積或平均離子或電子數目的儀器,它通常被用於某些涉及到偵測微弱的離子或電子訊號的實驗。以本實驗的研究課題為例,當離子被加速並打在微通道片(microchannel
plate)上,離子的數目即被放大約4
x 107倍且轉換成電流,利用50
Ohm的電纜線配上BNC接頭即可輸入多通道定標平均器,若電壓大於所設定的discrimination
level(例如-10 mV),則可被計數。實驗進行中可設定取樣時間(例如30
seconds),由多通道定標平均器可得知單位時間內所偵測到的總離子數目,或平均離子數目。以Stanford
Research Systems Model SR 430 Multichannel Scaler/Averager為例,它具有下列特性: Singal
Input
| Bandwidth |
dc to 250 MHz |
| Input
impedance |
50 Ohms |
| Linear range |
+/- 300 mV |
| Input
protection |
+/- 5 Vdc, 50
V for 1 microsecond |
Discriminator
| Discriminator
range |
-300 mV to
+300 mV |
| Discriminator
slope |
Rising or
falling |
| Resolution |
0.2 mV |
| Input offset
voltage |
1 mV |
| Minimum pulse
input |
10 mV |
| Pulse pair
resolution |
5 ns |
| DISC output |
NIM levels
into 50 Ohms |
Trigger
Input
| Impedance |
10 ohms |
| Threshold |
-2 V to 2 V |
| Slope |
Rising or Falling |
| Protection |
15 VDC |
Internal
Time Bins
| Bin width |
5, 40, 80, 160, ... ns, 10.486 ms |
| Accuracy |
1 ns |
| Jitter |
100 ps + 10 ppm of delay |
| Accuracy |
2 ns +1% |
Counter/Accumulation
| Bins per record |
1k to 16k in 1k increment |
| Max count rate |
100 MHz |
| Max count |
32,767 per bin per trigger |
| Records per accumulation |
1 to 64k (or free run) |
Trigger
Rate
Minimum period between
triggers
Tp = record time + accumulation time + overhead
= (no. of bins x bin width) + (no. of bins x 250
ns)
+ 150 microseconds Outputs
DISC = Nim level into 50 ohms
|